


Next:Comparison
Between Measured &
Up:High-Energy
Quantum Efficiency from
Previous:A
typical example and
Results for the flight devices.
The technique described above was applied to all frontside illuminated
flight devices. Depletion depths were extracted from the Fe55data
sets taken at the HIREFS spectrometer chamber. These data have lower flux
rate than other chambers, and, hence, very low level of pileup. The computation
of the depletion depth included correction for the presence of the
line in the source spectrum as well as correction for the undetected photons
according to equation (4.40). No pileup
correction was made because the level of pileup in these data is low. Results
are summarized in Table 4.30.
Table 4.30: Depletion Depth Estimated from 5.9 keV branching
ratio
| Device |
Location |
Depletion Depth,  |
Remarks |
PI = +2,+10V (standard) |
PI = -5,+5V (reduced dark current) |
|
| c0 |
c1 |
c2 |
c3 |
c0 |
c1 |
c2 |
c3 |
|
| w203c4r |
I0 |
66.4 |
66.9 |
66.8 |
66.4 |
47.6 |
48.1 |
48.0 |
48.7 |
|
| w193c2 |
I1 |
67.4 |
66.5 |
67.8 |
66.6 |
48.3 |
47.4 |
47.5 |
47.9 |
|
| w158c4r |
I2 |
66.8 |
67.2 |
66.5 |
66.3 |
47.7 |
47.9 |
47.8 |
47.4 |
|
| w215c2r |
I3 |
66.1 |
67.3 |
66.5 |
66.2 |
48.7 |
47.8 |
48.5 |
47.9 |
|
| w168c4r |
S0 |
64.8 |
65.5 |
65.5 |
65.3 |
44.9 |
45.8 |
45.5 |
45.4 |
|
| w182c4r |
S2 |
79.5 |
79.1 |
80.6 |
79.9 |
57.7 |
58.2 |
58.1 |
58.3 |
|
| w457c4 |
S4 |
73.3 |
74.2 |
73.9 |
73.6 |
53.4 |
53.5 |
53.0 |
53.4 |
|
| w201c3r |
S5 |
73.0 |
74.4 |
74.7 |
74.4 |
52.1 |
53.4 |
52.9 |
53.2 |
|
| w198c1 |
ref. |
73.7 |
73.9 |
74.7 |
72.8 |
52.3 |
51.6 |
52.2 |
51.3 |
|
| w461c4 |
ref. |
81.2 |
81.3 |
80.2 |
80.6 |
59.2 |
58.9 |
58.2 |
58.4 |
|
| w203c2 |
ref. |
68.1 |
68.1 |
67.0 |
67.2 |
48.0 |
49.0 |
48.9 |
47.8 |
|
|
For standard image section voltages (+2,+10 Volts) every measurement
was made at three different temperatures, namely, -130, -120, and -110
degrees C. Since depletion depths should not depend on temperature (at
least in the narrow temperature range), for every quadrant of every chip
the results were averaged over three measurements. Typical standard deviation
of the result is 0.35 microns. As can be seen in Table 4.30,
some of the chips show quadrant-to-quadrant variations of the depletion
depth on the order of 1.5 microns. Such variations exceed statistical error
and, most likely, are real and caused by nonuniformity of the substrate
resistivity. They translate into approximately 2% variations of QE at 8.4
keV and provide an explanation why spatial variation of QE at this energy
is much larger than at lower energies.



Next:Comparison
Between Measured &
Up:High-Energy
Quantum Efficiency from
Previous:A
typical example and
Please address comments and questions to Dr. John Nousek ( nousek@astro.psu.edu
)