Effective Area and Energy Resolution
vs. Split Event Threshold
The HRMA/ACIS effective area and energy resolution were also evaluated
for split event thresholds of 5 to 15 adu in steps of 2 adu.
Figure 6.5 and figure 6.6
show the dependence of effective area with split event threshold for ACIS
CCD's I1 and S3. Tabular versions of the data are listed in Tables 6.13
and 6.14. Figures 6.7
and 6.8 show the dependence
of energy resolution with split event threshold for ACIS CCD's I1 and S3.
Tabular versions of the data are listed in Tables 6.15
and 6.16.
These results indicate that the HRMA/ACIS effective area and energy
resolution do not vary considerably with split event threshold for FI CCD's,
whereas the BI chip performance is significantly different, showing a significant
dependence on split event threshold.
Figure 6.5: Effective Area
vs. Energy for ACIS CCD I1 calculated for split event thresholds of 5,
7, 9, 11, 13 and 15 ADU
Table 6.13: Effective Area vs. Split Threshold
for CCD I1
Energy
keV |
Effective Area |
| 5adu |
7adu |
9adu |
11adu |
13adu |
15adu |
| cm2 |
cm2 |
cm2 |
cm2 |
cm2 |
cm2 |
| 0.277 |
2.37 |
2.38 |
2.38 |
2.38 |
2.38 |
2.38 |
| 0.525 |
118.42 |
118.84 |
118.92 |
118.90 |
118.87 |
118.88 |
| 0.93 |
334.65 |
334.73 |
334.84 |
334.80 |
334.75 |
334.78 |
| 1.4867 |
569.80 |
570.06 |
570.10 |
570.06 |
570.06 |
570.03 |
| 3 |
284.07 |
286.06 |
286.15 |
286.31 |
286.28 |
286.3 |
| 4 |
339.75 |
344.87 |
345.65 |
345.87 |
345.91 |
346.14 |
| 5 |
305.37 |
315.73 |
318.24 |
319.62 |
320.80 |
321.60 |
| 6 |
137.45 |
153.44 |
161.07 |
165.56 |
168.88 |
171.71 |
| 7 |
52.08 |
63.97 |
71.07 |
75.96 |
79.82 |
82.86 |
| 8 |
19.66 |
25.06 |
28.67 |
31.63 |
34.23 |
36.31 |
| 9 |
8.12 |
10.19 |
11.75 |
13.13 |
14.44 |
15.63 |
The effective area values listed in this table apply for CCD
ID = 1, a HRMA Pitch angle of 5 arcmin, a HRMA Yaw angle of -3.7 arcmin,
and grade selection of g02346. Events within the entire CCD were considered.
Figure 6.6: Effective Area
vs. Energy for ACIS CCD S3 calculated for split event thresholds of 5,
7, 9, 11, 13 and 15 ADU
Table 6.14: Effective Area vs. Split Threshold
for CCD S3
Energy
keV |
Effective Area |
| 5adu |
7adu |
9adu |
11adu |
13adu |
15adu |
| cm2 |
cm2 |
cm2 |
cm2 |
cm2 |
cm2 |
| 0.277 |
137.02 |
159.79 |
165.75 |
167.60 |
168.43 |
169.10 |
| 0.525 |
295.19 |
360.90 |
384.35 |
391.97 |
394.43 |
395.79 |
| 0.93 |
439.99 |
535.67 |
572.61 |
586.07 |
591.09 |
592.62 |
| 3 |
167.70 |
245.80 |
299.84 |
332.17 |
348.93 |
355.37 |
| 4 |
124.12 |
189.12 |
241.82 |
282.05 |
307.53 |
321.12 |
| 5 |
76.47 |
124.05 |
165.86 |
199.18 |
224.27 |
240.84 |
| 6 |
45.15 |
69.54 |
92.15 |
112.64 |
129.18 |
140.04 |
| 7 |
35.28 |
50.68 |
63.98 |
76.14 |
87.31 |
94.09 |
| 8 |
16.70 |
22.91 |
28.42 |
33.20 |
37.13 |
39.94 |
| 9 |
10.29 |
14.86 |
18.51 |
21.87 |
24.54 |
26.58 |
The effective area values listed in this table apply for CCD
ID = 7, a HRMA Pitch angle of 0 arcmin, a HRMA Yaw angle of 0 arcmin, and
grade selection of g02346. Events within the entire CCD were considered.
Figure 6.7: Energy Resolution
(FWHM) vs. Energy for ACIS CCD I1 calculated for split event thresholds
of 5, 7, 9, 11, 13 and 15 ADU
Table 6.15: Energy Resolution vs. Split
Threshold for CCD I1
Energy
keV |
Energy Resolution |
| 5adu |
7adu |
9adu |
11adu |
13adu |
15adu |
| eV |
eV |
eV |
eV |
eV |
eV |
| 0.277 |
45.97 |
46.30 |
44.86 |
45.33 |
45.68 |
44.85 |
| 0.525 |
55.59 |
55.42 |
54.64 |
54.64 |
54.73 |
55.19 |
| 0.93 |
60.38 |
60.25 |
60.83 |
60.58 |
60.98 |
61.22 |
| 1.4867 |
77.29 |
77.43 |
77.61 |
77.57 |
77.79 |
77.92 |
| 3 |
91.21 |
90.05 |
90.60 |
90.25 |
91.82 |
91.00 |
| 4 |
103.57 |
103.24 |
102.99 |
103.16 |
103.27 |
103.87 |
| 5 |
116.07 |
115.49 |
115.20 |
116.24 |
116.43 |
117.01 |
| 6 |
129.83 |
131.17 |
131.70 |
132.58 |
134.95 |
135.17 |
| 7 |
147.54 |
151.02 |
151.19 |
152.15 |
151.91 |
153.10 |
| 8 |
150.85 |
155.08 |
159.14 |
158.46 |
164.41 |
169.73 |
| 9 |
178.35 |
186.48 |
183.48 |
184.22 |
189.48 |
193.95 |
The energy resolution values listed in this table apply for
CCD ID = 1, a HRMA Pitch angle of 5 arcmin, a HRMA Yaw angle of -3.7 arcmin,
and grade selection of g02346. Events within the entire CCD were considered.
Figure 6.8: Energy Resolution
(FWHM) vs. Energy for ACIS CCD S3 calculated for split event thresholds
of 5, 7, 9, 11, 13 and 15 ADU
Table 6.16: Energy Resolution vs. Split
Threshold for CCD S3
Energy
keV |
Energy Resolution |
| 5adu |
7adu |
9adu |
11adu |
13adu |
15adu |
| eV |
eV |
eV |
eV |
eV |
eV |
| 0.277 |
43.35 |
62.63 |
62.57 |
66.68 |
70.07 |
76.64 |
| 0.525 |
83.09 |
80.60 |
82.10 |
86.62 |
92.34 |
98.31 |
| 0.93 |
99.03 |
96.89 |
95.71 |
98.30 |
102.10 |
108.68 |
| 3 |
130.55 |
135.66 |
139.76 |
139.35 |
137.55 |
138.21 |
| 4 |
146.38 |
150.33 |
152.01 |
150.78 |
149.48 |
150.50 |
| 5 |
162.78 |
171.01 |
175.40 |
176.75 |
173.58 |
171.24 |
| 6 |
167.08 |
177.24 |
182.56 |
186.36 |
184.27 |
182.22 |
| 7 |
178.56 |
191.37 |
195.14 |
195.39 |
193.79 |
186.53 |
| 8 |
192.94 |
195.28 |
202.41 |
206.32 |
205.97 |
199.89 |
| 9 |
172.92 |
187.21 |
195.52 |
204.09 |
207.44 |
207.28 |
The energy resolution values listed in this table apply for
CCD ID = 7, a HRMA Pitch angle of 0 arcmin, a HRMA Yaw angle of 0 arcmin,
and grade selection of g02346. Events within the entire CCD were considered.