X-ray Line Flux Density at HRMA Entrance

As described in section 6.2.1 an absolute effective area measurement requires a normalization to the X-ray flux density at the HRMA entrance. For DCM measurements in the 3 to 10 keV range the X-ray beam incident on the HRMA was measured to be non-uniform at approximately the 10% level. We have estimated the incident flux by taking the average of the four BND's at the HRMA location. Higher spectral orders from the DCM are not expected to be present in the 3 to 10 keV range. The quantum efficiency values and open areas for the BND counters were provided by SAO.

For the present reduction of the BND spectra we have chosen a region of interest approach. The extracted X-ray line count rate is divided by the quantum efficiency of the BND at the X-ray line energy.