The high flux levels obtainable (
1
1012 photons/sec/0.1% BW) on this beamline were comparable to
the levels obtained at NSLS. Additionally, the stability of the source
beam, the high spectral purity and wavelength accuracy at the ALS allowed
us to obtain high quality transmission data which clearly resolve the near
and extended absorption fine structure of the Al-L, C-K, N-K and O-K absorption
edges (see Figures 5.6 - 5.9).
As an indicator of the absolute accuracy of the transmission values
a comparison of the ALS and NSLS transmission values away from absorption
edges indicates differences of less than 1% of the measured value. Figure
5.10 shows the X-ray transmission
function of the ACIS-I and S filters between 0.05 and 3 keV. Transmission
values below 0.7 keV are from the ALS measurements, transmission values
between 0.7 and 1.8 keV are from the NSLS measurements.
![]() |
We have created a transmission function of the ACIS OBF's between the energies 0.05-10 keV as a composite of measurements and models. (Fig 5.10) In particular in the energy range between 0.05 - 0.7 keV we use the X-ray transmission values derived from the measurements of the OBF's at the ALS at Berkeley Laboratories, in the energy range 0.7 - 1.8 keV we use the transmission values obtained at the NSLS at Brookhaven Laboratories and between 1.8 - 10 keV we have used a model that incorporates results from fits to the measured transmission data.