next up previous contents
Next: Effects of CCD Clocking Up: Temperature Dependence on Dark Previous: Post Cold Soak Characterizations

Summary of Measurements

For future reference, we present the dark current data taken at the ISIM TV test in Table 4.68. Included is the dark current (ADU/sec) and the width (ADU) for each chip.


 
Table 4.68: Dark Current Measured at ISIM TV tests during October, 1997. In each column, the number to the left if the dark current (ADU/sec), the number to the right is the width (ADU) of the gaussian fit.
  T=-40 $ \mbox{$^{\circ}$ }$ C T=-60 $ \mbox{$^{\circ}$ }$ C T=-90 $ \mbox{$^{\circ}$ }$ C
Chip     pre cold soak post cold soak
I0 22.9 10.4 1.13 1.63 -0.05 1.03 -0.14 1.13
I1 24.1 10.1 1.37 1.72 -0.01 1.06 -0.07 1.10
I2 24.7 10.2 1.41 1.76 0.01 0.96 -0.02 1.09
I3 22.9 8.9 1.04 1.61 -0.06 1.01 -0.12 1.20
S0 23.6 10.0 0.99 1.68 -0.04 1.08 0.02 1.29
S2 18.1 8.3 1.04 1.50 -0.06 0.95 -0.02 1.29
S4 18.1 7.9 1.06 1.66 0.01 1.29 0.08 1.40
S5 22.5 13.4 1.14 1.76 -0.10 1.17 -0.04 1.21
S1 65.0 68.3 2.79 4.04 -0.03 1.25 -0.06 1.43
S3 55.6 54.0 1.76 3.25 0.06 1.35 -0.08 1.20
 


next up previous contents
Next: Effects of CCD Clocking Up: Temperature Dependence on Dark Previous: Post Cold Soak Characterizations
Please address comments and questions to Dr. John Nousek ( nousek@astro.psu.edu )