nextuppreviouscontents
Next:Uncertainties in the Cross-calibration Up:Relative Quantum Efficiency Measured Previous:Relative Quantum Efficiency Measured

  
Spatially-averaged Relative Quantum Efficiency Measurements

The results of the relative quantum efficiency measurements described above in Section 4.7.1 are presented in Table 4.52. The mean relative efficiencies given in the table are the centroids obtained from Gaussian fits to histograms of the superpixel ratios in each measurement. These values have been corrected for pileup and are our best estimates for the spatially averaged relative efficiency between the corresponding flight and reference detectors. Listed in parentheses are measurements of the spatial variation (RMS) in efficiency across each chip, obtained by computing the standard deviation of the superpixel ratios. (Note that the RMS variations in efficiency are heavily influenced the framestore cover shielding at the edge of each device. An alternative measure of spatial variation, which is more representative of the majority of the area of each device, is presented and discussed in Figures 4.77-4.86 in section 4.7.2.2.) The name of the reference device used during each measurement is listed in the table, along with the focal plane position of each flight device. Note the QE ratio listed for w140c4r at 525 eV is referenced to w134c4r, a backside chip. No measurement relative to w103c4 was made at this energy because of time constraints. Also, the QE ratio for w134c4r relative to w203c2 at 677 eV was derived from the corresponding ratio for w140c4r relative to w134c4r. This was necessary because reference device w203c2 was misaligned during calibration of w134c4r at this energy.
 
 
Table 4.52: Spatially Averaged Relative Detection Efficiency at CSR
 
Flight Position Mean (RMS) Relative Efficiency vs. Energy (keV) Reference
Device   0.525 0.677 1.740 2.015 4.509 5.894 8.040 Device
    O F Si P Ti Fe55 Cu  
w203c4r I0 0.972 0.992 0.996 0.982 0.994 0.977 0.933 w190c3
    (0.040) (0.041) (0.046) (0.046) (0.046) (0.045) (0.043)  
w193c2 I1 0.885 0.768 0.980 0.951 1.005 1.034 1.097 w103c4
    (0.045) (0.040) (0.052) (0.049) (0.054) (0.055) (0.060)  
w158c4r I2 0.914 0.804 0.986 0.968 1.007 1.029 1.096 w103c4
    (0.040) (0.037) (0.047) (0.045) (0.047) (0.048) (0.055)  
w215c2r I3 0.999 1.015 0.997 0.989 0.996 0.980 0.929 w190c3
    (0.041) (0.040) (0.051) (0.050) (0.052) (0.053) (0.046)  
w168c4r S0 1.064 1.089 1.004 1.014 0.997 0.972 0.916 w190c3
    (0.042) (0.044) (0.050) (0.050) (0.050) (0.048) (0.047)  
w182c4r S2 1.009 0.962 1.003 1.002 1.014 1.067 1.210 w103c4
    (0.042) (0.042) (0.053) (0.052) (0.053) (0.055) (0.064)  
w457c4 S4 1.020 1.070 1.005 1.020 1.000 0.999 1.000 w190c3
    (0.030) (0.043) (0.031) (0.032) (0.031) (0.031) (0.031)  
w201c3r S5 1.004 1.025 1.001 0.991 1.000 1.000 1.005 w190c3
    (0.032) (0.034) (0.038) (0.037) (0.039) (0.039) (0.040)  
w140c4r S1 1.030 1.078           w134c4r
    (0.037) (0.038)            
      2.507 1.035 1.442 0.934 0.763 0.631 w103c4
      (0.114) (0.050) (0.082) (0.080) (0.054) (0.041)  
w134c4r S3 3.457 2.955 1.108 1.570 1.014 0.882 0.743 w203c2
    (0.113) (0.184) (0.061) (0.087) (0.061) (0.054) (0.051)  
w190c3 Ref. 0.890 0.767 0.986 0.949 1.010 1.058 1.164 w103c4
    (0.020) (0.017) (0.018) (0.014) (0.014) (0.016) (0.051)  
w203c2 Ref. 0.872 0.787 0.979 0.950 1.004 1.009 1.041 w103c4
    (0.016) (0.016) (0.015) (0.015) (0.015) (0.016) (0.021)  

The correction factors used to correct the count rates at CSR for pileup are presented in Table 4.53. The values listed in parenthesis are the pileup correction factors applied to the count rates from the reference devices.
 
 

Table 4.53: Pileup Correction Factors Applied to Count Rates from CSR
 
Flight Position Flight (Reference) Pileup Correction vs. Energy (keV) Reference
Device   0.525 0.677 1.740 2.015 4.509 5.894 8.040 Device
    O F Si P Ti Fe55 Cu  
w203c4r I0 1.053 1.074 1.075 1.097 1.063 1.068 1.169 w190c3
    (1.055) (1.075) (1.076) (1.100) (1.064) (1.070) (1.185)  
w193c2 I1 1.061 1.031 1.033 1.033 1.026 1.064 1.067 w103c4
    (1.071) (1.041) (1.076) (1.079) (1.060) (1.146) (1.142)  
w158c4r I2 1.068 1.032 1.022 1.033 1.027 1.066 1.066 w103c4
    (1.076) (1.041) (1.050) (1.078) (1.061) (1.151) (1.141)  
w215c2r I3 1.051 1.075 1.073 1.095 1.066 1.068 1.166 w190c3
    (1.051) (1.075) (1.074) (1.097) (1.067) (1.070) (1.182)  
w168c4r S0 1.068 1.074 1.073 1.097 1.067 1.070 1.166 w190c3
    (1.064) (1.068) (1.074) (1.096) (1.068) (1.073) (1.184)  
w182c4r S2 1.042 1.039 1.033 1.034 1.026 1.065 1.073 w103c4
    (1.042) (1.041) (1.075) (1.078) (1.059) (1.143) (1.140)  
w457c4 S4 1.067 1.074 1.078 1.098 1.066 1.073 1.200 w190c3
    (1.066) (1.069) (1.078) (1.096) (1.066) (1.074) (1.201)  
w201c3r S5 1.065 1.070 1.078 1.096 1.065 1.072 1.184 w190c3
    (1.065) (1.069) (1.079) (1.098) (1.066) (1.073) (1.184)  
w140c4r S1 1.073 1.037           w134c4r
    (1.072) (1.034)            
      1.037 1.030 1.048 1.032 1.049 1.018 w103c4
      (1.040) (1.076) (1.078) (1.061) (1.140) (1.144)  
w134c4r S3 1.063 1.033 1.068 1.096 1.077 1.057 1.057 w203c2
    (1.049) (1.031) (1.070) (1.062) (1.060) (1.062) (1.176)  
w190c3 Ref. 1.062 1.070 1.067 1.091 1.052 1.078 1.087 w103c4
    (1.070) (1.093) (1.069) (1.097) (1.119) (1.172) (1.166)  
w203c2 Ref. 1.035 1.031 1.034 1.032 1.026 1.056 1.070 w103c4
    (1.041) (1.040) (1.078) (1.076) (1.059) (1.127) (1.158)  

In Table 4.54 we have used the values given in Table 4.52 to compute QE ratios referenced to w190c3 for each flight device. The results are plotted in Figure 4.76. Most of the relative efficiency points in the upper panel of the figure lie within 5% of unity, and all but two of the points lie within 10%, indicating that at each energy the absolute efficiencies of the eight frontside devices are nearly identical. In the lower panel we see the backside devices are about three times more efficient than the frontside devices at low energies (< 1 keV), but are less efficient than the frontside devices at high energies (> 3-4 keV).
 
 

Table 4.54: Spatially Averaged Relative Detection Efficiency at CSR--Referenced to w190c3
 
Flight Position Mean (RMS) Relative Efficiency vs. Energy (keV)
Device   0.525 0.677 1.740 2.015 4.509 5.894 8.040
    O F Si P Ti Fe55 Cu
w203c4r I0 0.972 0.992 0.996 0.982 0.994 0.977 0.933
    (0.040) (0.041) (0.046) (0.046) (0.046) (0.045) (0.043)
w193c2 I1 0.994 1.002 0.994 1.002 0.996 0.977 0.942
    (0.053) (0.054) (0.056) (0.053) (0.055) (0.054) (0.076)
w158c4r I2 1.027 1.049 1.000 1.020 0.997 0.972 0.941
    (0.048) (0.050) (0.050) (0.050) (0.049) (0.049) (0.073)
w215c2r I3 0.999 1.015 0.997 0.989 0.996 0.980 0.929
    (0.041) (0.040) (0.051) (0.050) (0.052) (0.053) (0.046)
w168c4r S0 1.064 1.089 1.004 1.014 0.997 0.972 0.916
    (0.042) (0.044) (0.050) (0.050) (0.050) (0.048) (0.047)
w182c4r S2 1.133 1.255 1.017 1.056 1.005 1.008 1.039
    (0.052) (0.058) (0.057) (0.056) (0.055) (0.055) (0.083)
w457c4 S4 1.020 1.070 1.005 1.020 1.000 0.999 1.000
    (0.030) (0.043) (0.031) (0.032) (0.031) (0.031) (0.031)
w201c3r S5 1.004 1.025 1.001 0.991 1.000 1.000 1.005
    (0.032) (0.034) (0.038) (0.037) (0.039) (0.039) (0.040)
w140c4r S1 3.487 3.269 1.050 1.519 0.925 0.721 0.542
    (0.190) (0.154) (0.054) (0.088) (0.080) (0.053) (0.047)
w134c4r S3 3.386 3.032 1.100 1.572 1.008 0.841 0.664
    (0.139) (0.203) (0.065) (0.093) (0.064) (0.055) (0.062)

 

Figure 4.76:  Relative detection efficiencies of front-illuminated (upper panel) and back-illuminated (lower panel) ACIS flight CCDs as a function of energy - referenced to w190c3.

We have computed the mean relative efficiency and the maximum absolute deviation from the mean for each CCD type as a function of energy (see Table 4.55). The largest absolute deviations from the mean ($\sim$10-20%) in the frontside devices are at 525 eV, 677 eV, and 8 keV. The deviations at 525 eV and 677 eV are due to the high relative efficiency of w182c4r at these energies, which in turn is caused by that detector's relatively thin oxide layer. The reason for the large deviations at 8 keV is still unknown.

From the RMS variations listed in Table 4.54 it can be seen that spatial variations in efficiency across the frontside devices are low ($\sim$3-6%). About two-thirds of the RMS variation is the result of shadowing by the framestore covers. Thus, we find the detection efficiencies of the frontside devices are highly uniform within each chip and between the chips.

The spatial variations in efficiency across the backside devices are generally larger ($\sim$5-8%) than in the frontside devices, but the maximum absolute deviations from the mean of the backside devices are of comparable amplitude to those of the frontside devices. Hence the spatially averaged quantum efficiencies of the two backside devices are in reasonably good agreement with each other.
 
 

Table 4.55: Mean Relative Detection Efficiencies for FI and BI CCDs at CSR--Referenced to w190c3
 
Device Mean (Max. Abs. Dev.) Relative Efficiency vs. Energy (keV)
Type 0.525 0.677 1.740 2.015 4.509 5.894 8.040
  O F Si P Ti Fe55 Cu
FI 1.027 1.062 1.002 1.009 0.998 0.986 0.963
  (0.106) (0.193) (0.015) (0.046) (0.006) (0.022) (0.076)
BI 3.436 3.151 1.075 1.546 0.967 0.781 0.603
  (0.050) (0.119) (0.025) (0.027) (0.041) (0.060) (0.061)


nextuppreviouscontents
Next:Uncertainties in the Cross-calibration Up:Relative Quantum Efficiency Measured Previous:Relative Quantum Efficiency Measured
Please address comments and questions to Dr. John Nousek ( nousek@astro.psu.edu )